"mpa and its metabolites were analysed as follows:
C18 column (4.6 x 150 mm, 5µ ) was eluted with an water/methanol mobile phase, where the concentration of methanol was linearly increased from 52 to 80% methanol over 20 min. The flow rate was set at 1 ml/min and the UV detection wavelength was set at 254 nm.
The eluent was introduced into the MS instrument with a split ratio of 1:4. The mass spectrometer was a TSQ triple quadrupole (Thermo Fisher Scientific, Waltham, MA) equipped with an electrospray ionization interface. The spray voltage was set at 4.5 kV, and the capillary temperature was 300°C. The auxiliary and nebulizing gas was Nitrogen. The nebulizing gas backpressure was set at 40 psi, and auxiliary gas was set at 20 (arbitrary units). A full MS scans between m/z 200 and 500 was performed to observe the molecular ions."

Analytical Test

Identification

Analytical Technique

LC

References

  1. Zhang, Jiang-Wei; Liu, Yong; Zhao, Jiu-Yang; Wang, Li-Ming; Ge, Guang-Bo; Gao, Yang; Li, Wei; Liu, Hong-Tao; Liu, Hui-Xin; Zhang, Yan-Yan; Sun, Jie; Yang, Ling. Metabolic profiling and cytochrome P450 reaction phenotyping of medroxyprogesterone acetate. Drug Metabolism and Disposition. 2008; 36:2292 - 2298